Remote type electromagnetic flowmeters are widely used in industrial applications. Their design allows the transmitter and flow sensor to be installed separately, making them suitable for high temperature, high pressure or hazardous environments. However, some common problems may be encountered during use. The following are remote type electromagnetic flowmeter common issues and solutions. When encountering related problems, you can refer to the table below.
Troubles & Solutions
Common Issues | Possible Causes | Solutions |
---|---|---|
Unstable output | Unstable flow field | Modify the pipeline or add a dummy sensor |
The liquid passing through the sensor contains gas or large solid blocks | Normal phenomenon | |
Electrical connection is loose | Check the wiring and connect the wires properly | |
Poor grounding | Connect the ground wire properly | |
Electrode leakage | Repair the sensor | |
No output for liquid flow | The two core wires of the signal transmission cable between the sensor and the converter are reversed | Reverse the wire head |
Power supply is not connected or poor contact | Connect the power supply and ensure good contact | |
Leakage from the sensor instrument pipe, shell, or end face | Repair the sensor | |
There is output when the liquid is not flowing | Open circuit in the signal transmission cable connection between the sensor and the converter | Connect the cable properly |
Open circuit in the signal cable connection to the electrode | Open the sensor and reconnect it | |
Contamination or insulation layer deposition on the electrode surface | Clean the electrode surface | |
Poor grounding or open circuit | Connect the ground wire properly | |
Excessive error | Zero point is too high | Readjust the zero point |
Not fully filled with liquid | Improve the pipeline conditions, the sensor is always filled with liquid | |
Excessive distortion of the power supply | Improve the power supply conditions to meet normal operating conditions | |
Poor grounding | Connect the ground wire properly |
The above are common issues, possible causes and solutions for remote type electromagnetic flowmeters. If your flowmeter encounters the above problems during use, you can refer to the solution, or you can contact us at any time, we will do our best to troubleshoot for you to restore use as soon as possible.