4 different types of water analysis controllers

Water Analysis Controller Can Monitor Different Parameters

Water analysis controller is a device used for real-time monitoring and control of water quality parameters, widely used in the fields of water treatment, environmental protection, industrial production, etc. It acquires various parameter data from water samples through sensors, such as pH value, dissolved oxygen, turbidity, conductivity, etc., and performs data analysis and processing to ensure that the water quality meets the predetermined standards. Choosing the right water analysis controller can significantly improve the efficiency and reliability of water treatment systems.

  • Specification
  • Video
Specification
  • A universal water analysis controller

    SUP-DC2000 Universal Controller

  • A pH6.0 water analysis controller

    SUP-PH6.0 Controller

  • A pH water analysis controller

    SUP-PH (CCEP) Controller

  • A conductivity water analysis controller

    SUP-TDS210-B Conductivity Controller

  • A turbidity water analysis controller

    SUP-PTU300 Turbidity Controller

  • A multi-parameter water analysis controller

    SUP-MDX500 Multi-parameter Controller

  • A residual chlorine water analysis controller

    SUP-TRC400 Residual Chlorine Controller

SUP-DC2000 universal controller, is suitable for supporting use with various liquid analysis series digital sensors of our company. It is used to monitor parameters including pH, ORP, conductivity, dissolved oxygen, turbidity sludge concentration, etc. The monitored parameters are output to the monitoring room through RS485 or current transmission for record keeping.

Features

  • Sensors can be calibrated offline
  • Automatic sensor recognition
  • Support various parameters, pH/ORP, conductivity, dissolved oxygen, turbidity, residual chlorine, etc.
A universal water analysis controller

Parameters

Table 1: SUP-DC2000 Universal Controller Parameters
Display 2.8-inch LCD display
Range pH: 0–14, ORP: -2000mV to +2000mV
Dissolved oxygen: 0–20 mg/L / 0–40 mg/L
Saturation: 0–200%
Conductivity: 0–600 ms/cm
Turbidity: 0–4000 NTU
SS/TSS: 0–120,000 mg/L
Relay output One set of high and low limit alarms (3A250V AC), normally open contact relay
Communication RS-485 communication, MODBUS RTU
Transmission Isolated 4–20mA output, the maximum loop is 750Ω, ±0.2% F.S
Overall dimension 100 mm × 100 mm × 150 mm
Cutout dimension 92.5+1 mm × 92.5+1 mm

Parameters

Table 2: SUP-PH6.0 Controller Parameters
Display 2.8-inch LCD display
Range pH: 0–14
ORP: -1000mV to +1000mV / -2000mV to +2000mV
Relay output One set of high and low limit alarms (3A250V AC), normally open contact relay
Communication RS-485 communication, MODBUS RTU
Transmission Isolated 4–20mA output, the maximum loop is 750Ω, ±0.2% F.S
Overall dimension 100 mm × 100 mm × 150 mm
Cutout dimension 92.5+1 mm × 92.5+1 mm

Parameters

Table 3: SUP-PH (CCEP) controller Parameters
Display 4.3-inch TFT display
Range pH: 0–14
ORP: -1999mV to +1999mV
Relay output One set of high and low limit alarms (3A250V AC), normally open contact relay
Communication RS-485 communication, MODBUS RTU
Transmission Isolated 4–20mA output, the maximum loop is 750Ω, ±0.2% F.S
Overall dimension 144 mm × 144 mm × 115 mm
Cutout dimension 138+1 mm × 138+1 mm

Parameters

Table 4: SUP-TDS210-B Conductivity controller Parameters
Display 2.8-inch LCD display
Range 0.01 electrode: 0.20–200.0 μS/cm
0.1 electrode: 2.00–2000 μS/cm
1.0 electrode : 0.02–20.0 mS/cm
10.0 electrode: 0.20–200.0 mS/cm
Temperature: -10 °C to +130 °C
Relay output the relay load is 3A/250V AC
Communication RS-485 communication, MODBUS RTU
Transmission Isolated 4–20mA output, the maximum loop is 750Ω, ±0.2% F.S
Overall dimension 100 mm × 100 mm × 150 mm
Cutout dimension 92.5+1 mm × 92.5+1 mm

SUP-PTU300 turbidity controller is for online monitoring of drinking water quality,widely used in on-line monitoring of turbidity in tap water factory water, secondary water supply, membrane filtration water, swimming pools, surface water, etc. It has the characteristics of ultra-low turbidity detection limit, long-term-maintenance-free high-precision measuring equipment, water-saving work and digital output.

Features

  • Suitable for low turbidity conditions
  • Innovative integrated body, sophisticated structure design
  • Wall-mounted installation, easy and convenient
  • Third-generation laser light source technology, without external measurement probe
  • Less manual maintenance, fast detection
A turbidity water analysis controller

Parameters

Table 5: SUP-PTU300 Turbidity Controller Parameters
Dimension 66 mm × 145 mm × 54 mm
Ingress protection IP54
Measuring range 0–1 NTU, 0–20 NTU, 0–100 NTU (optional)
Accuracy ±2% or ±0.015 NTU (Based on formazin primary standard at 25 °C)
Operating temperature 0°C – 50 °C
Storage temperature -20 °C to 60 °C
Zero point drift ≤ ±0.015 NTU
Resolution 0.001 NTU
Water inlet and outlet Water inlet 6mm hose, outlet and sewage outlet 10mm hose
Inlet flow 50–300 mL/min
Sensor cable length 2 m
Installation Wall-mounted

MDX500 multi-parameter water quality online analyzer is a new generation drinking water quality monitoring equipment independently developed and manufactured by our company. This equipment can be widely used in urban or village waterworks, tap water pipeline network, tap water secondary water supply, user terminal, indoor swimming pool, large water purification equipment and direct drinking water and other water quality online monitoring.

Features

  • Simultaneously monitoring multiple parameters such as turbidity, pH, temperature, etc.
  • High-precision measurement, long-term stable and accurate measurement on the order of 0.001–0.1 NTU and 0.1–1 NTU
  • Wall-mounted installation to prevent water flooding and ground moisture, convenient for installation and operation and maintenance
  • With a self-protection device, can effectively avoid equipment damage caused by accidents and lightning strikes
  • Low operation and maintenance costs, support remote control functions such as automatic sewage discharge and remote adjustment
  • Strong environmental adaptability, optional temperature control heating antifreeze module, can run outdoors all year round in cold regions
A multi-parameter water analysis controller

Parameters

Table 6: SUP-MDX500 Multi-Parameter Controller Parameters
Parameters Turbidity Chlorine/Chlorine Dioxide pH/ORP (optional) Temperature Conductivity (optional) Dissolved oxygen (optional)
Measuring range 0–1 NTU / 0–20 NTU
0–100 NTU / 0-4,000 NTU
0–5 mg/L / 0–20 mg/L pH: 0–14
ORP: ±2000mV
0–50 °C 0–2000 μS/cm 0–20 mg/L
Resolution 0–1 NTU / 0–20 NTU
0.001 NTU: 0–100 NTU
0.01 NTU: 0–4,000 NTU
0.01 mg/L pH: 0.01
ORP: ±1mV
0.1 °C
Lower detection limit 0.02 NTU
0.1 NTU: 0–4,000 NTU
0.05 mg/L
Zero point drift ≤ 1.5%
Indication stability ≤ 1.5%
Accuracy 2% o r± 0.02 NTU
2% or 0.1 NTU (0–4,000 NTU)
±0.05 mg/L or ±5% (DPD comparison error ±10%) pH: ±0.1
ORP: ±20mV or ±2
±0.5 °C ±1.5% FS ±0.3 mg/L
Repeatability ≤ 3% pH: 0.1
ORP: ±10mV
≤ 0.5 °C ≤ 0.5% FS ≤ ±1.5%
Response time T≤ 120 s, Measured value: 0–90% of turbidity value ≤ 120 s ≤ 60 s ≤ 25 s ≤ 30 s ≤ 30 s
Recommended maintenance period 3–12 months (depending on the water quality on site) 1–3 months or weekly calibration
3–6months to replace consumables
1–3 months 12 months 3–6 months 1–3 months

Parameters

Table 7: SUP-TRC400 Residual Chlorine Controller Parameters
Display 7-inch touch screen
Weight 8 kg
Ingress protection IP43
Range Residual chlorine: 0–5 mg/L
Temperature: 0.1–40 °C
Accuracy
Resolution 0.01
Zero point drift
Power supply (220±22)V AC, (50±1) Hz
Operating temperature 0–40 °C (No condensation)
Relative humidity ≤ 95% (No condensation)
Inlet flow ≥ 0.03 m/s (in the flow cell)